Page 32 - Plastics News April 2019
P. 32

Thermogravimetric Analysis (TGA)
     In order to compare the degradation profile of the two samples, thermogravimetric analysis (TGA) was carried out using a
     TA Instruments Q500 instrument.  Approximately 15 mg of sample was placed in a tared platinum TGA pan and heated
     from 30°C to 900°C at 10°C per minute in an atmosphere of air.  The inset Figure shows the TGA curves that were
     recorded for this pair of samples.
     The degradation profiles and the amount of inorganic residue were compared between the two samples.  No significant
     differences were noted between the two samples.
     A different thermal degradation profile would have indicated a substantially different polymer or additive package.  If a
     different degradation profile had been observed, chemical identification of the base polymer followed by an analysis of the
     stabilizer package would have been performed.  Analytical methods employed would include Fourier transform infrared
     (FTIR) spectroscopy and nuclear magnetic resonance (NMR) spectroscopy.
     A difference in inorganic residue content would have indicated a difference in the amount or type of filler that was
                  FeAtures
     compounded into the thermoplastic elastomer resin.  If a significant difference in inorganic residue content had been
     observed then the chemical composition, concentration, and the physical characteristics of the filler would have been
     documented.  Scanning electron microscopy (SEM) coupled with energy dispersive X-ray spectroscopy (EDS) provides
     this additional data, together with digital optical microscopy (OM).


                                                                heated a second time to 250°C at a heating rate of 10°C
                                                                per minute. The Figure 3 shows the DSC curves that were
                                                                recorded for this pair of samples. Both samples show two
                                                                glass transitions on the first heating scan that vary slightly.
                                                                Both thermoplastic polyurethanes also demonstrate a
                                                                melting point at 155°C. In addition to having the same
                                                                melting peak characteristics (peak onset, peak minimum,
                                                                peak termination) the level of crystallinity, as determined
                                                                by the area of the melting peak (heat of fusion, ΔHf), is
                                                                equivalent for the two samples.
                                                                Conclusions
                                                                Melt flow index, thermogravimetric analysis, and
                                                                differential scanning calorimetry were used to determine

                                Fig.2                           the root cause of substantial process variation with
                                                                an extruded thermoplastic polyurethane. TGA testing
          Fourier transform infrared (FTIR) spectroscopy and nuclear   revealed  no  significant  differences  between  Good  and
                                                                Bad samples in terms of the thermal degradation profile
          magnetic resonance (NMR) spectroscopy. A difference
          in inorganic residue content would have indicated a   or inorganic residue contents. No significant difference in
          difference in the amount or type of filler that was   either of the two glass transition temperatures or in the
                                Page 2 of 3
          compounded into the thermoplastic elastomer resin.    detailed characteristics of the melting endotherm were
          If  a  significant  difference  in  inorganic  residue  content   noted either. The most marked dissimilarity between the
          had  been  observed  then  the  chemical  composition,   two samples was in the melt flow index (MFI) value. The
          concentration,  and  the  physical  characteristics  of  the   melt flow rate was about twice as high for the Bad sample
          filler would have been documented. Scanning electron   relative to the Good sample. These significantly different
          microscopy (SEM) coupled with energy dispersive X-ray   melt flow rate values indicate a large variation in the
          spectroscopy (EDS) provides this additional data, together   molecular weight distributions for these two thermoplastic
          with digital optical microscopy (OM).                 polyurethane samples.(Courtesy: SGS Polymer solutions)
  Differential Scanning Calorimetry (DSC)
  DSC was performed using a TA Instruments Q200.  Approximately 7 mg of sample was encapsulated in an aluminum pan
          Differential Scanning Calorimetry (DSC) DSC was performed
                                                                 The Importance Of MFI
  and heated from -80°C to 250°C at a heating rate of 10°C per minute. The samples were then cooled to -80°C and
          using a TA Instruments Q200. Approximately 7 mg of sample
  subsequently heated a second time to 250°C at a heating rate of 10°C per minute.  The inset Figure shows the DSC
  curves that were recorded for this pair of samples.
          was encapsulated in an aluminum pan and heated from
  Both samples show two glass transitions on the first heating scan that vary slightly.  Both thermoplastic polyurethanes
                                                                 The
  also demonstrate a melting point at 155°C.  In addition to having the same melting peak characteristics (peak onset, peak  melt flow index (MFI) is a measure of the ease
          -80°C to 250°C at a heating rate of 10°C per minute.
  minimum, peak termination) the level of crystallinity, as determined by the area of the melting peak (heat of fusion, ΔH f), is
                                                                 of flow of the melt of a thermoplastic polymer. It is
          The samples were then cooled to -80°C and subsequently
  equivalent for the two samples.
                                                                 defined as the mass of polymer, in grams, flowing in
                                                                 ten minutes through a capillary of a specific diameter
                                                                 and length by a pressure applied via prescribed
                                                                 alternative gravimetric weights for alternative
                                                                 prescribed temperatures. Melt flow rate is an indirect
                                                                 measure of molecular weight, with high melt flow rate
                                                                 corresponding to low molecular weight. At the same
                                                                 time, melt flow rate is a measure of the ability of the
                                                                 material's melt to flow under pressure. Melt flow rate
                                                                 is inversely proportional to viscosity of the melt at the
                                                                 conditions of the test, though it should be borne in
                                                                 mind that the viscosity for any such material depends
                                                                 on the applied force. Ratios between two melt flow
                                                                 rate values for one material at different gravimetric
                                                                 weights are often used as a measure for the broadness
                                                                 of the molecular weight distribution.
   Conclusions                  Fig.3
   Melt flow index, thermogravimetric analysis, and differential scanning
   calorimetry were used to determine the root cause of substantial process
   variation with an extruded thermoplastic polyurethane.  TGA testing revealed
                                     32
          Plastics News  April  2019
   no significant differences between Good and Bad samples in terms of the
   thermal degradation profile or inorganic residue contents.  No significant
   difference in either of the two glass transition temperatures or in the detailed
   characteristics of the melting endotherm were noted either.  The most marked
   dissimilarity between the two samples was in the melt flow index (MFI) value.
   The melt flow rate was about twice as high for the Bad sample relative to the
   Good sample.  These significantly different melt flow rate values indicate a
   large variation in the molecular weight distributions for these two
   thermoplastic polyurethane samples.






                                 Page 3 of 3
   27   28   29   30   31   32   33   34   35   36   37